Fisica generale e sanitaria Modulo di: Imaging e fisica sanitaria Ä M MOLIS Ä SCIEN RUN MAL MCMLXXXI REI Ing. Giovanni Mastrangelo Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni MastrangeloDIAGNOSTIC MEDICAL IMAGING 3rd Part - Nuclear Magnetic Resonance Imaging Ä M MOLIS Ä SCIEN RUN MAL MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 2
Physics of Magnetic Resonance Magnetic resonance scanners use the property of nuclear magnetic Resonance (NMR) to create images All nuclei have positive charges (they are composed by protons and neutrons). A nucleus with either an odd atomic number or an odd mass number has an angular momentum - they have spin The nuclei of the hydrogen atoms (1H) have spin 1/2 Rotating charges create an angular magnetic momentum, the rotating nuclei of the hydrogen atoms similar to little magnets N + + Nucleus angular momentum Microscopic magnetization of nucleus + + + + + S M MOLL Ä Ä CIEN Ä RUA MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 3
Physics of Magnetic Resonance In normal conditions individual spins of 1H nuclei have a random orientation, → → has results no macroscopic magnetic field is produced If the nuclei of the hydrogen atoms (1H) are subjected to a strong magnetic Field, B0 , they tend to align with the field (parallel and anti-parallel orientation); being the number of hydrogen atoms into the human body very high, this tendency results in a magnetization of the body A little more than half of the nuclei will be oriented in the same direction of the external magnetic field, "up" oreintation (having a lower energy content), while the others nuclei will be oriented in the opposite direction, "down" oreintation (having higher energy content). Random thermodynamic interaction between magnetic dipoles and the surrounding macro-molecules create a continuous change in spin orientation. The dynamic difference between "up" and "down" nuclei is strictly related the external field intensity. M MOLIS Ä RUN MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 4
Physics of Magnetic Resonance Macroscopic magnetisation grows as B0 grows A magnetic field of 1.5 T creates a difference between up and down spin oreintation of 6 p.p.m. 1 mmc of tissue 1019 hydrogen nuclei a significant magnetic magnetisation is produced Actually, nuclei spin precess around an axis along the direction of the field. This precession has a frequency, called Larmor frequency (proportional to B0 ), of the order of MHz (radiofrequency). As an example the precession frequency for an external field of 1.5 T is 64 MHz Bo M MOLIS E Ä SCIEN Ä RUA MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 5
Macroscopic Magnetisation It is important to underline that there is no precession phase coherence between nuclei The xy plane components of the microscopic magnetisation vectors are placed in every directions, their sum is zero. This component of the macroscopic magnetisation is called Transversal Macroscopic Magnetisation TMM0=0 The z plane components of the microscopic magnetisation vectors are placed both in z and -z directions. Their sum is positive along z direction. This component of the macroscopic magnetisation is called Longitudinal Macroscopic Magnetisation LMM0±0 Z LMM #0 X y TMMg=0 y X IM MOLIS Ä CIE Ä Ä RUN MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 6
Macroscopic Magnetisation The Macroscopic Magnetisation produced by the external magnetic field is static and smaller than B0 it is not possible to measure it An external excitation has to be introduced in the system If a microscopic sample of nuclei is excited using a electromagnetic radiation having Larmor frequency, the radiation magnetic component interacts with nuclei magnetic moment A quantum of energy is absorbed changing the nuclei energy status from “up" to "down" The longer is the RF pulse duration the higher is the energy transfered to the system (the higher is the number of nuclei that changes their status from up to down) When these energy transitions occur, nuclei are resonant with applied radiation The Bloch equations are a set of coupled differential equations which can be used to describe the behavior of a MM vector M MOLIS E CIEN Ä RUA Ovy MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 7
Macroscopic Magnetisation When the difference between up and down spin populations decreases their precession phase syncronisation grows It is possible to set the RF pulse duration to reach the condition of having the same "up" and "down" spin population that have complete precession phase coherence This RF is called "90° Pulse" The system subject to the 90° Pulse has LMM1=0 while the TMM reach its maximum, the TMM rotate in the xy plane with the nuclei precession frequency (the RF pulse one) The MM, following a spiral movement with growing ray, deflects from its position along z axis to a position in the xy plane, so it is rotated by 90° TMM, growing y Ä CIEN RUA PUB MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 8 Z X y LMM, decreasing
Macroscopic Magnetisation If the duration of the RF pulse is doubled (w.r.t. the one of the 90° pulse) It is possible to obtain the total "up" and "down" populations reversal (w.r.t. the initial situation), in this status there is no precession phase coherence between nuclei LMM2= - LMM., TMM2=0 This RF is called "180° Pulse" It is clear that it is possible to change RF pulse amplitude and duration so that different degrees of deflection of MM can be obtained The displacement of MM from its longitudinal direction leads to a unstable system (no energetic equilibrium), having high energy and high phase nuclear synchronisation When the external RF ends the system goes back to initial conditions IM MOLIS Ä Ä RUA MAL MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 9
Relaxation When the external RF ends, the rapidly rotating (and decreasing) TMM creates a radio frequency excitation within the sample that will in turn induce (Faraday induction) a voltage in a coil of wire located outside the sample. This signal is recorded for use in MRI. Transverse relaxation, also known as spin-spin relaxation, acts first to cause this received signal to decay. This relaxation is caused by the perturbations in the magnetic field due to other spins that are nearby. This interaction causes spins to momentarily speed up or slow down, changing their phases relative to other nearby spins (dephasing). The resulting signal is called free induction decay (FID), an exponential decay having a time constant called transverse relaxation time, T2 Ä M MOLIS Ä SCIEN RUN MAL MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 10
Free Induction Decay FID amplitude is a function of the number of nuclei in the sample, the decreasing time is a function of the decreasing speed of TMM 1.0 - 0.5 - 0.0 Time 0 -0.5- -10 It is not easy to obotain a good FID sampling, due to the fact that the transmitting and receiving coils are the same, moreover the FID signal decreases very quickly and is effected by non homogeneous magnetic field of the structure The RF pulses practically used in NMR imaging create spin echoes E M MOLI E Ä SCIEN Ä RUN MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 11 Receiver Coil Signal Amplitude x-y plane
Relaxation T2 is the time required to the TMM to reach the 37% of its initial value (reached when the RF pulse is applied) There are two factors causing the decreasing of phase coherence of resonant nuclei -The energy transitions of nuclei that changes their status from down to up - Interactions between spin and molecular micro-EM field, these interactions do not vary the energy of the system but create local variations of the static field,B0 ,inducing a change in precession velocity (phase displacement) Actually, local perturbations of B0cause the received signal to decay exponentially with a time constant T2*, lower than T2 This effect is reversible using the technique of coherence refocusing thorugh the concept of echoes (that will be described later) M MOLIS Ä Ä RUN MAL MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 12
Relaxation The second relaxation mechanism is called longitudinal relaxation or spin-lattice relaxation.This process concerns the longitudinal magnetisation which recovers back to its equilibrium value as a rising exponential having a time constant called longitudinal relaxation time, T1 T1 is the time required to the LMM to reach the 63% of its equilibrium value T1 is a function of how fast the spin systems release energy to return from down to up position, this thermodynamic energy exchange involves the surrounding molecules (this process is a function of B0 ) T1 and T2 are different for various types of tissues and are responsible for generating contrast in MR images. For tissue in the body the relaxation times are in the ranges: 250 ms & T1 ≤ 2500 ms , 25 ms ≤ T2 ≤ 250 ms Usually 5T2 ≤ T1 ≤ 10T2 and for all materials T2 ≤ T1 This is due to the fact that there are spin-spin interactions that cause phase coherence loss without changing up and down populations therefore T1 relaxation gives contribution to T2 relaxation but not vice-versa IM MOLIS Ä Ä SCIEN RUA MCMLXXXI REI Imaging e fisica sanitaria A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 13
Relaxation Mz = Mo ( 1 - e-t/T1 ) Mz M. M.(1-e1) T1 t T, constants (in ms) 0.2 Tesla 1.0 Tesla 1.5 Tesla Fat 240 Muscle 370 730 863 White matter 388 680 783 Gray matter 492 809 917 CSF 1,400 2,500 3,000 IM MOLL E Ä IEN RUA MAL MCMLXXXI REI Imaging e fisica sanitaria MxY =MxYo e-t/T2 My e T2 t T2 constants (in ms) Fat 84 Muscle 47 White matter 92 Gray matter 101 CSF 1,400 A.A. 2023/2024 Ing. Giovanni Mastrangelo Pagina 14